In this paper, finite element analyses for estimating the behavior of fractured femur just after the operation were carried out by using ABAQUS 6.71. A stainless steel bone plate and composite bone plates with various stacking angles were considered to find out the effect of bone plate properties on bone fracture healing. In order to simulate the actual state, contact conditions between the plate and bone and fractured bones were imposed on the finite element models and the whole analysis was divided by two steps; screw fastening step and load bearing step. The stress and strain distributions at the fracture site for the cases of the stainless steel and composite bone plates were analyzed and compared with. From the analyses it was found that the composite bone plate had potential advantages for effective bone fractures healing relieving stress shielding effect.
본 논문에서는 상용 유한요소해석 프로그램인 ABAQUS 6.71을 이용하여 수술 직후 골절부 거동을 예측할 수 있는 해석을 수행하였다. 스테인리스 고정판과 적층순서를 달리한 복합재료 고정판을 대상으로 대퇴골의 골절치료에 대한 효과를 비교 분석 하였다. 실제 상태를 모사하기 위해 고정판과 골절뼈 및 골절부간 접촉조건을 부여하였으며, 스크류 체결과정과 하중부가 과정을 두 단계로 나누어 해석을 수행하였다. 유한요소해석을 통하여 스테인리스 고정판과 복합재료 고정판이 체결된 경우에 대한 골절부 틈새의 변화 및 응력분포 등을 계산하고 그 차이를 비교하였다. 복합재료 고정판은 골절뼈와의 접촉에 의한 불필요한 응력 증가를 감소시키고 골절부위의 응력은 증가시켜 골절치료에 더 효과적인 것으로 나타났다.
Keywords: Finite Element Analysis, Contact condition, Bone plate, Femur
Keywords: 유한요소해석, 접촉조건, 고정판, 대퇴골